Recently Updated Pages

备份

三星XR

开机,在提示按F12 选USB Storage Debice

Updated 3 hours ago by Admin

新页面

企业

 

Updated 3 weeks ago by Admin

图像质量

飞利浦MR指南

开始>Service Application Planned Maintenance>SPT Batch Files>PIQT>鼠标右键>Run Batch Batch Files>IQT>鼠标...

Updated 4 weeks ago by Admin

诊断

飞利浦MR指南

开始>Service Application Corrective Matntenance>Diagnostics> Spike Detemination>Spile test Body col...

Updated 4 weeks ago by Admin

查看水冷

飞利浦MR指南

开始>Service Application Corrective Matntenance>Diagnostics>Liquid Cooling Cabinet(LCC)>LCC Info ov...

Updated 4 weeks ago by Admin

用户密码

飞利浦MR指南

MR 账户   mruser 无 mrservice manager Philips123

Updated 4 weeks ago by Admin

日志

飞利浦CT指南

D:\Data_Logger\Logger.mdb   C:\usr\diamond.root\bin\Service Tools\ViewLogsPage.htm    

Updated 1 month ago by Admin

注册

企业

Updated 1 month ago by Admin

创办

企业

经营范围 经营范围规范表述查询系统(https://jyfwyun.com/home) 一般项目:互联网销售(除销售需要许可的商品);单用途商业预付卡代理销售;软件开发;软件销售;数字技术服务;...

Updated 1 month ago by Admin

GE DR6000 球管面板触摸屏失灵,需要重装系统

GE医疗XR案例

Problem Description Details 球管面板触摸屏失灵 Problem Cause Problem Solution 需要重新安装系统,步骤如下 1. 用光盘先备份系统设置和...

Updated 2 months ago by Admin

GE CT 机架电源灯亮机架部分没有电

GE医疗CT案例

Problem Description Details 机架没有电 Problem Cause 机架部分使用不了 Problem Solution 1;电源柜电源指示灯是亮的机架部分没有电 2:...

Updated 2 months ago by Admin

GE CT 床异响

GE医疗CT案例

Problem Description Details 设备升降床液压杆下降和上升有异响 Problem Cause 设备升降床异响 Problem Solution 1:升降床液压杆里面有异响...

Updated 2 months ago by Admin

GE CT Optima 安装选件-EKG 心电监护

GE医疗CT案例

Problem Description Details 安装选项 Problem Cause 安装选件 Problem Solution 1、放入 EKF 选件光盘到操作台上光驱,进入维修界面找...

Updated 2 months ago by Admin

GE CT Optima680 做No Load KV 测试

GE医疗CT案例

Problem Description Details 高压部分或者球管故障 Problem Cause 高压部分或者球管故障 Problem Solution 1、断开球管高压电缆做 No L...

Updated 2 months ago by Admin

GE CT Optima 680 球管旋阳故障及更换球管

GE医疗CT案例

Problem Description Details 设备扫描报错40-0114,Rotor not at speed。 Problem Cause 1、高压电缆故障。 2、球管故障。 3、H...

Updated 2 months ago by Admin

GE CT Optima 680 报错60-0313

GE医疗CT案例

Problem Description Details 设备报:硬件扫描部分停止扫描,设备无法使用。 报错 60-0313:inverter max. ILR current detected。...

Updated 2 months ago by Admin

GE CT 添加PACS

GE医疗CT案例

Problem Description Details GE CT 添加PACS Problem Cause ‐‐ Problem Solution 点击工具按钮进入维修界面,如图: 点击App...

Updated 2 months ago by Admin

GE revolution ACTs 做no load KV

GE医疗CT案例

Problem Description Details 扫描报错开始曝光后检测到的KV 太低、608h Problem Cause 球管打火 Problem Solution 1、在机器待机时候...

Updated 2 months ago by Admin

校准

赛德科XR指南

球管朝向地面,拉最低,关闭曙光器 点击左上角电脑图标, 关闭系统>取消传输全部数据>恢复模式 在Task Manager>Startup>CXDI>Disabled 重新计算机 机架关机断电,打...

Updated 2 months ago by Admin

检查方案

飞利浦XR指南

系统>设置EVA>启动 用户名Application specialist 密码extended EVA>Database>部位>视图 选择View parameters>Acuisiton d...

Updated 2 months ago by Admin